RESPIREASY dedakj DE-2A Home Oxygen Concentrator Oxygen Therapy Instructions


RESPIREASY HOME OXYGEN CONCENTRATOR
USER MANAUL
BRAND INTRODUCTION
RESPIREASYOxygen Concentrator is designed and manufactured by Shenzhen Jiaxing Technology Innovation Co., Ltd.
RESPIREASY, developed by Shenzhen Jiaxing Technology Innovation Co., Ltd., has specialized in oxygen concentrators for over 15 years. We provide 1–60 LPM portable, home, and medical-grade devices with original accessories, serving 400,000+ users in 60+ countries.
Germany R&D Based & Shenzhen Factory
Backed by German respiratory experts in Germany R&D Based, we hold 100+ patents and focus on PSA efficiency, AI adaptive flow, and IoT monitoring. Our ISO 13485-certified Shenzhen factory (CE/ISO/NMPA) has a 2M annual capacity, with strict 7-step testing ensuring 93% purity and <0.5% failure rate in 2 years.
RESPIREASY delivers reliable, affordable, and smart oxygen therapy solutions for home, travel, and clinical care.
--RESPIREASY, easy respiratory care, smart oxygen therapy solutions

CONTENTS
Safety precautions---------------------------------- 2
Product features------------------------------------ 3
Key technical indexes------------------------------- 3
Installation and usage------------------------------ 4
Care and maintenance------------------------------ 8
Common trouble-shooting------------------------- 10
Operation principle--------------------------------- 12
In case of power outage and malfunction of theoxygen concentrator,an alternate source of supplemental oxygen (such as oxygen cylinder or oxygen bag)should be prepared for the user in urgent need of oxygen.The machine is not intended as medical device and is not tobe used for life support or life extension.
Upside down operation is prohibited.
f the power supply has unstable voltage, or is beyond AC220+22(110+10)volt, please connect the voltage regulator before using the machine.Please use standard socket with electrical safety qualifcation and terminal board with electricalsafety certifcation.
The machine should be located in a well-ventilated space and avoid direct sunlight. The distance between the machine and the wall or other objectsshould be no less than 10cm. Do not put other things or liquid vessel on themachine.
There should not be any obstruction around the airoutlet on the back and bottom of the machine,incase of air inlet and air outlet blocked, which couldlead to high temperature that will result in machine stoppage and decreasing of oxygen concentration.
Oxygen is a combustion-supporting gas, so do not smoke and keep away from combustion source while using the machine for fear of fre accident. Do not switch on & off frequently. You can restart it after you have switched it off for 3-5 minutes, considering its service life. Avoid oil or grease-like material for any part of the machine. So as not to pollute the oxygen machine and cause fre.
The scope of product application: The product is designed in accordance with standard of Home personalusage appliance.
Intelligent control: LED display screen, easy use and operations operate for timing function.Overheat protection: the pure oil-free air compressor with overheat protector, will stop working automatically when the compressor's working temperature is too high to protect the compressor and the whole part of the product.
This product uses AC220V/110V as power source, air as raw material and adopts the pure oil-free air compressor and qualifed molecular sieve imported with original packaging, so as to generate high purity oxygen by PSAat normal temperate and supply oxygen.
Product name: Oxygen concentrator
Oxygen generation method: Pressure Swing
Rated voltage: AC220+22V 50+1Hz/AC110+10V 60Hz
Oxygen flow rate: 2L-9L/min adjustable
Oxygen pressure:0.02-0.05MPa
Operating noise:≤45dB(A)
Classifcation:class II
Size:404mmx182mmx383mm
Net weight: 8.0kgDisplay control: LED light display, Intelligent chip control, infrared remote control
Open-package inspection



Filling water in humidifcation water tank: Grip the top handle of the humidifcation water tank by hand,pull out the water tank after pressing




LED screen displayinstruction
6.lnfrared remote control

(1)On/0f key
(2)Timing + key
(3)Timing-key
(4)Negative ion key
(5)Sleep mode key
In working state, press "Sleep" mode key, the "Sleep" iconon the display screen will fash three times,then the screen will turn into sleep mode. Press any key on the remote control or touch any key on the machine panel to active the screen highlight.
Operational temperature:5°C-40°C Relative humidity range:≤80%RH Atmospheric pressure range: 86kPa-106kPa
When the temperature is below 5°C,the machine should be put in normal operating condition for at least 4 hours before getting started.
Transport or storage conditions Storage temperature:-20°C-60°C Relative humidity range:≤93%RH
To avoid strong electromagnetic interference environment or interfering other domestic appliances, please keep the machine at least 3m away from other domestic appliances.
1. Cleaning housing unit
To keep the water tank clean,the water used should be purifed water, distilled water or cold boiled water.we suggest to change water every day.
5. Long-term storage operations
Avoid direct sunlight and pollution. Preserve the oxygen tube and oxygen inhaler at room temperature. Do not twist the airway tube. Keep out of the reach of the children.
Common Trouble-shooting
| NO. | Phenomenon | Cause analysis | Solution |
| 1 | Odor in air. | In the initial stage of using, there is smell for the connection tube and other plastic parts. It is normal. | The odor that comes from silicone tube and ABS rubber parts is normal and non-toxic by inspection. The odor will disappear as time goes by. |
| The humidifier water tank hasn't been cleaned for a long time or the water hasn't been changed. | Clean the oxygen tube and humidifier water tank. | ||
| 2 | Water flows from the oxygen outlet. | The water level of the humidifier water tank is beyond the highest water level, so the airflow pressed the bubble into the oxygen tube. | Keep the water lower than the highest water level. |
| The machine has been used for a long time, so that the steam in the airflow gathers on the tube wall. | Pour the water in water tank. And add water into the water tank when there is no water coming out from oxygen outlet. | ||
| 3 | The machine lets "puff" sound every several seconds | It is the sound of nitrogen discharging which is normal and is a sign of fine working. | |
| 4 | After startup, the machine operates well, the flowmeter displays "0" or is too small | The flowmeter is off or the value is too small. | Turn up the oxygen flow. |
| The oxygen tube is twisted or blocked. | Straighten the oxygen tube. | ||
| The inner connection tube of the machine is disconnected or broken. | Disconnect the power cord and contact professionals to repair it. |
| 5 | After startup, there is bubble in the water tank, but there is no oxygen getting out from the oxygen outlet. | The silica gel stopper of the humidifying water tank is not properly positioned. | Remove the humidifying water tank to plug the silica gel stopper. |
| 6 | After startup, the indicator light works, but the machine does not (the sound is irregular). | Compressor malfunction. | Disconnect the power cord and contact professionals to repair it. |
| The temperature is so high that the self-protection mode of the oil-free compressor is activated. | It is a normal case. When the temperature lowers down, the machine will restart automatically. | ||
| 7 | Separation valve malfunction. | Disconnect the power cord and contact professionals to repair it. | |
| The inner connection tube of the machine is disconnected or broken. | Disconnect the power cord and contact professionals to repair it. |
The air comes into the compressor after compressed by air inlet flter unit and then it enters into molecular sieve tower to separate the oxygen and nitrogen. The oxygen successfully goes through the molecular sieve tower into the fne sieve tower. In contrast, the nitrogen is absorbed by the molecular, then to be discharged into the atmosphere through the separating valve. The oxygen concentration is enhanced in the fne sieve tower and controlled fow size by fowmeter. Then the oxygen is humidifed by the humidifying water tank when it comes to it. Finally, the oxygen reached the user through oxygen supply tube.