Portable Oxygen Concentrator Instructions KPA

Portable Home Oxygen Concentrator
INSTRUCTION
Instruction and Technical Manual
Please read the manual thoroughly before use.
The pictures are for reference only and the machine in kind shall prevail.

Contents
Product Diagram                              
Product Specifications                               
Interface Introduction                                
Directions For Use                                      
Maintenance And Cleaning                   
Warranty Description                              
Common Problems And Solutions        
Packing List                                                  
Alarm Information                                          
Repair Information 

Product Namel:portable oxygen concentrator KPA
Product model:See the body label for details
Product size:L*W*H :17.0cm*8.0cm*19.8cm
Product weight:Machine weight :1.45Kg ,Battery weight :0.45Kg
Interface:2.8"monitor
Operating noise:≥45dB
 Audio alarm:≤60dB( A )( position 5 )
Sea level:5800M normal use
Warm-up time:2min
oxygen concentration:≥90%( At all gears and respiratory rates )
Gear setting:1 2 3 4 5 6 7 8 (different  models and different gears )
Respiratory rate:10-40 BPM
Respiratory sensitivity:≤0.12 cmH2O
Max outlet pressure:100KPa±20%
AC Power Source:AC100-240V 50/60 Hz 60VA MAX
Internal battery:14.4VDC 4A Max
Battery duration:8-cell battery:4~1.2h (Varies according to the ambient temperature)
Battery charging time:8-cell battery: 3.5h
Suitable environment:Temperature ,5~40℃.Humidity ,10%~90%. non condensing state .Atmospheric pressure ,540hPa~1060hPa.

The above data are tested by the laboratory, and the final interpretation right belongs to the company

Settings Interface

Long press ON/ OFF
Click on the sensitivity button ,to select the appropriate sensitivity.
Click on on the timing button, to select timing.
Click the plus  or minus,button to select the appropriate flow.

DIRECTIONS FOR USE

1. Power supply
Select the appropriate power connection conditions based on the usage environment.
When only using batteries, install the dedicated battery for this device to the device battery interface.
When the vehicle power adapter is used, if the power indicator on the adapter is on, it in
dicates that the power connection is normal.   Required input terminal: DC 19v , power100W.
When using the AC power adapter, connect the AC power adapter and the power cord firmly.
When the power indicator on the adapter lights up, it indicates that the power connection is normal.
Required input terminal: AC 100-240V, current 2.0-1.0A, and frequency 50/60Hz.
 Warning: Do not use lithium batteries that are not equipped with the product.
2. Startup preheating
Pressthe device ON/OFF button,and after  hearing a "beep" prompt tone for startup, enter the startup screen, and the device enters the preheating state  after the systemstarts.  The preheating time is about 2 minutes. Once preheating is complete and the device
reaches the normal operating state,  you can wear  the rubber hose and start normal use of  it.
   Note: During the machine preheating, the device will automatically spray oxygen to empty the internal air. At this time, the output flow and concentration of the
device may not meet the standard. Please do not use it.
   Note: Please do not connect the rubber hose and operate the machine during the machine preheating.
3. Connecting the rubber hose
Connect the flared end of the factory-supplied rubber hose to the device's oxygen outlet nozzle. Ensure the connection is secure and airtight. Make sure the rubber
hose is  not  twisted    or blocked to avoid triggering alarms and affecting  normal use.
   Note: This device is pulse oxygen supply mode, and the rubber hose must be connected when the device is used.
4. Adjusting the flow gear
Please press the set button to set the flow gear according to your own requirements.
5. Shutdown
When the device is working normally,   long press the ON/OFF button for about 3 seconds, the device  will beep and enter the shutdown screen, and the equipment will automatically shut
down.
6. Other information
When the adapter is connected or the battery is connected, the equipment will   beep and display the current power supply information on the equipment home screen.
When an abnormal state occurs, the abnormal information indicator light will light up, and the equipment will display theabnormal information and issue an audible prompt.
   Note: It is prohibited to remove the battery during equipment operation.
   Note: This  product  is  not  suitable  for  medical  use

MAINTENANCE AND CLEANING

1. Replacement of intake filter cotton
(1) Shut down the device, disconnect the power supply and remove the battery.
(2) Remove the filter cotton bottom shell.
(3) Take out the intake filter cotton. Note: The air intake filter cotton may accumulate a  large amount of dust after long-term use. Handle with care and take precautions to avoid inhaling dust.
(4) Install new filter cotton in the correct direction.
(5) Install the filter cotton bottom shell back.
(6) Please properly dispose of the waste filter cotton.
 Note: Failure to install or replace the filter cotton for a long time may affect equipment.performance or damage the equipment. Please replace the filter cotton in a timely manner, and the replacement cycle depends on the use environment and usage. Generally, it is replaced once a month.

2. Molecular sieve replacement and maintenance
(1) When the device prompts the need to replace the molecular sieve, please contact the designated seller in a timely manner.
(2) Long term storage may shorten the service life of molecular sieves.
(3) Long term operation in a humid environment may shorten the service life of molecular sieve.
 Warning: Do not disassemble and replace molecular sieves without permission. Only   equipment suppliers or qualified maintenance engineers can disassemble and replace molecular sieves.

3. Replacement of rubber hose
Users can purchase rubber hoses according to their needs, but they need to ensure that the following conditions are fully met:
(1) Please contact the equipment provider or purchase under the guidance of a professional.
(2) Please purchase from regular channels.
(3) The rated flow rate of oxygen in the rubber hose is 10L per minute, and the total length is 1-2m;
(4) It can fit well with the air outlet of this product without air leakage.
Please wear a rubber hose properly and use it according to the  instructions for using the rubber hose.

4. Battery replacement and maintenance
This device requires the use of a dedicated battery provided at the factory, with a standard
battery provided at the factory. Users can contact a designated seller to purchase batteries as needed.
Battery life: 500 cycles of charge and discharge.
When not in use for a long time, please remove the battery from the device and pay attention to protecting the battery electrodes from contacting conductors such as metals to prevent fire and other hazards.
Please store the battery away from children to avoid danger.
Batteries that have not been used for a long time need to be regularly charged and maintained. It is recommended that the battery power be kept at about 30% to 60% when stored.
  Warning: Do not use lithium batteries that are not equipped with the product.

5.Warranty description
Category :Oxygen concentrator host.  Molecular sieve.  Battery
 Description: Warranty, 1 years.  Warranty,  1 year.  500 times of full charge and discharge or 1 year warranty
  Note: The service life in the table is the recommended value, and the actual service life    will vary depending on the actual usage environment and actual usage. Please pay close attention to the equipment usage status and related alarm prompts.

6. Maintenance cycle
Name:  Intake filter cotton.  Molecular sieve assembly.   Battery
Maintenance cycle:  Each mont/100  hours.  once a year.  once three months
Maintenance method:  Userreplacement.   Paidreplacement.   Charge to 30%
Remarks:  Can be purchased separately.   Replace as prompted by the device.  No maintenance required for normal use
Note: The maintenance cycle in the table is the recommended value, and the actual
maintenance cycle will vary depending on the actual usage environment and the actual usage. Please pay close attention to the equipment usage and related alarm prompts.
 
7. Equipment cleaning
Before cleaning, please ensure that the machine is turned off, unplug the power supply from the socket, and remove the battery.
The power off sequence should be to disconnect the power plug on the wall first and then disconnect the plug on the machine side.
Do not let the equipment get wet or enter water. Once this happens, it will lead to equipment failure or shutdown, and it will also increase the risk of electric shock.
Do not pour liquid directly onto the machine. Chemical cleaning agents used on the plastic exterior of the machine may damage the plastic housing of the equipment. These cleaning  agents include, but are not limited to, the following lists: alcohol and alcohol products, high concentration chlorinated solutions (chloroethane), and oil products (Clorox).
Be careful not to allow liquid to enter the interior of the machine. Pay special attention to ensuring that the oxygen outlet is free of dust, water, or other particles.
Do not use organic solvents or other flammable and explosive volatile substances for cleaning. Please ensure that the equipment has been completely dried before use.
 Warning: Before disposal, remove batteries, etc., and contact relevant departments to properly dispose of equipment and accessories in accordance with local government  regulations.

WARRANTY DESCRIPTION

1. The oxygen concentrator  host (excluding consumables) is warranted for 1 year from the date of purchase.
2. All maintenance services must be carried out by a designated maintenance service center.
3. During the warranty period, faults not caused by human factors are warranted free of charge.
4. After the warranty period, maintenance costs are charged.
5. If damage is caused due to improper use, paid maintenance services shall be provided.
6. Consumable such as rubber hoses  and filter cotton are not covered by the warranty

COMMON PROBLEMS AND SOLUTIONS

problem:   
(1)Device cannot be opened.    
(2)No oxygen output
Possible causes: 
(1) The battery is not properly    installed or securely inserted.  
(2) Dead battery. 
(3) Poor AC power connection. 
(4) Equipment failure.  
(5) Device not turned on.   
(6)The rubber hose is kinked or blocked.   
(7)Equipment failure.
Recommended Solution:   
(1)Remove the battery and reinstall it correctly.
(2) Insert the battery and connect the adapter to charge the battery.(3)Check the power connection.
(4)If the above operation cannot solve the problem, please contact your equipment supplier.
(5)Turn on the oxygen generator
(6)Check the nasal oxygen tube and its connection to the oxygen outlet.
(7)Please contact the equipment supplier.

The following states are not faulty
The exhaust port is a heat dissipation port. When operating for a long time or the ambient temperature is high, the temperature of the gas discharged from the equipment will rise,   which is a normal phenomenon.

PACKING LIST

1- Oxygen concentrator host - Standard configuration
1- Battery - Standard configuration
1- AC adapter (with power cord) - Standard configuration
1- Rubber hose - Standard configuration
1- Intake filter cotton - Standard configuration(one built- in oxygen concentrator)
1- Portable package - Standard configuration
1- Instructions - Standard configuration
1- Warranty card - Standard configuration
1- Certificate - Standard configuration

ALARM INFORMATION

 

When a malfunction occurs,  you will hear  a "beep" sound as an alarm. At the
same time, alarm information will be displayed on the screen.  Please follow the instructions  below for troubleshooting.

Alarm Item:
(1) No breathing detected.
(2) Low battery.
(3) Compressor shutdown.
(4) System temperature is too high.
(5) Fan stopped.

Alarm reason:
(1) No breathing detected during normal operation   for 15 consecutive seconds.
(2) The battery power is between 0% and 20% when only using battery power.
(3) Compressor does not started or faulty.
(4) The system temperature is higher than 60°C.
(5) The fan stops rotating.
Alarm   Interval:  10s

Screen display and prompts:
(1) No breathing detected
(2) Low battery
(3) Compressor shutdown
(4) System temperature is too high
(5) Fan stopped

processing method:
(1) Please connect the rubber hose for use.
(2) Please charge the battery promptly.
(3) Check whether the machine is over heated, and whether the air inlet and cooling holes are blocked.Restart themachine after  10 minutes of shutdown. Pleasecontact the seller promptly.
(4) Check whether the air inlet and coolingholes of t he machine are blocked.  Restart the machine after 20 minutes of shutdown.
(5) Alarm only, please   contact the seller in a timely manner.

 

Quality and Service

Fast Delivery Time:

Within 3-4 days the parel can be delivered usually.

Now we are operating the USA/ Canada/ Mexico local warehouses.

Quality Warranty:

As reliable supplier, all oxygen machines in our store can enjoy 2 year quality warranty.

USA Mexico EU and UK Local Warehouse:

Yes, now we are operating the USA Mexico EU and UK Local Warehouses. The good quality, fast delivery and after-sales service can be guaranteed.

 

About Us

Respire Easy of Fresh Oxygen

Welcome to DEDAKJ- Respireasy Office Store, your trusted destination of 1 liter -20 liters home portable oxygen concentrators and DEDAKJ oxygen accessories replacements.

Airoxygenconcentrator is a store that focuses on oxygen concentrators and oxygen accessories. Our goal is to provide customers with affordable, high-quality air oxygen concentrators to maintain an active family lifestyle and improve their quality of life.

Top rate oxygen concentrators supplier

DEDAKJ- Respireasy is a professional oxygen concentrator manufacturer, has been deeply involved in producing oxygen concentrators O2 making machines for more than 14 years.
The 1-9 LPM home use oxygen concentrators, 10-20 LPM powerful O2 generator, 1.2 kgs portable mini size oxygen generators, original dedakj oxygen filter, humidifier water tank, remote control accessories parts for DE-1A, 1B, 1LW, ddt-1a 1b, DE-2A, 2AW, 1S, 2SW, C1L, Q1W DEDAKJ are offering.
Fast delivery time and 1 year quality warranty, stocks in the United States, Canada, Mexico and Europe. All oxygenerator prices include shipping cost.

 

 

 

Help in need? Feel free to contact us