                                                                                                   OPERATION MANUAL

                                             Medical Oxygen Therapy Concentrator Model:RE-M10L

BRAND INTRODUCTION 

RESPIREASY Oxygen Concentrator is designed and manufactured by Shenzhen Jiaxing Technology Innovation Co., Ltd.

RESPIREASY has specialized in oxygen concentrators for over 15 years. We provide 1–60 LPM portable, home, and medical-grade devices with original accessories, serving 400,000+ users in 60+ countries.

Germany R&D Based & Shenzhen Factory 

Backed by German respiratory experts in Germany R&D Based, we hold 100+ patents and focus on PSA efficiency, AI adaptive flow, and IoT monitoring. Our ISO 13485-certified Shenzhen factory (CE/ISO/NMPA) has a 2M annual capacity, with strict 7-step testing ensuring 93% purity and <0.5% failure rate in 2 years.

RESPIREASY delivers reliable, affordable, and smart oxygen therapy solutions for home, travel, and clinical care.

--RESPIREASY, easy respiratory care, smart oxygen therapy solutions

                              Contents

1 Foreword-------------------------------------------2
2 Symbols--------------------------------------------2
3 Safety notice---------------------------------------3
4 Product introduction------------------------------4
5 Operation condition-------------------------------5
6 Scope of application-------------------------------5
7 Technical parameter--------------------------------5
8 Structures and Functions---------------------------6
9 Operation instructions-----------------------------8
10 Alarms and safety devices------------------------11
11Maintenance---------------------------------------12
12 Troubleshooting----------------------------------14
13 EMC Declaration----------------------------------16
14 Accessories----------------------------------------19
15 Condition for transportation and storage--------20
16 Quality warranty-----------------------------------20
17 Contact us-----------------------------------------20

                                                                         1


1  Foreword

Thank you for purchasing our products, hoping you will be satisfied with our products.
This operation manual contains function, operation steps, basic trouble solution and so
on.
To ensure your efficient use of the oxygen concentrator, please have a close read of this operation manual before operating it.

2  Symbols

The following table is a list of symbols and definitions that used with the RespirEasy RE-M10L Oxygen Concentrator.

 

Symbol

Description

Symbol

Description

 

Warning - Describes a hazard or unsafe practice that 

can result in severe bodily injury or death

 

(∈0197        

Indicating its conformity with the

Medical Devices Directive

93/42/EEC.The 0197 is the number of the Notified Body.


Caution - Describes a hazard or unsafe practice 

that can result in property  damage

 

"ON"(power)

 

Follow User's Manual

 

"OFF"(power)

 

CLASS Il equipment

SN

Serial  number

 

Authorised Representative in the European Community

 

Separate collection for electrical and electronic equipment

 

 

Type BF Applied Part,

F-TYPE APPLIED PART complying

with the specified requirements of this standard to 

provide a higher degree of protection against electric

 shock than that provided by TYPE B APPLIED

PARTS.The symbol of Type BF

Applied Part will be pasted on the outlet of oxygen.

 

Variability, rotational adjustment. To identify the control

 by means of which a quantity is controlled. The

controlled quantity

increases/decreases by rotation

with the figure width

 

Date of manufacture

 

Alternating current

Manufacturer

F6.3AL/250V

Type and rating of fuse

 

Fragile, handle with care

       

Keep dry

 

This way up

 

Stacking limit by number

 

No open flame; Fire, open ignition source and smoking prohibited

       

No smoking

                                                                            2

3 Safety notice

  1. This unit is not a life-support device , and in certain circumstances oxygen therapy can be hazardous, it is suggested that if any patient who needs oxygen treatment, please follow doctor's advice to choose the right flow and period for oxygen before using the oxygen concentrator.
  2. The device was expected to use in a health care environment and only trained Medical staff can operate the device
  3. In the event of an alarm, you observe your oxygen concentrator not working properly, or if you feel discomfort, consult your Equipment Provider and /or your physician immediately.
  4. Use only voltage specified on rating label.
  5. This device manufactures high concentration oxygen, which promotes rapid burning. Keep oxygen concentrator far away from open flames and no smoking around the patient.
  6. Do not leave a nasal oxygen cannula under bed coverings or chair cushions. If the unit is turned on without use, the oxygen will help the flammable material get fire. :Use no lubricants, grease, or petroleum-based products on or near your oxygen concentrator.
  7. Electrical shock hazard. Do not remove covers while the unit is plugged in. Only your Equipment Provider or a qualified service technician should remove the covers or service the unit.
  8. Care should be taken to prevent the unit from getting wet or allowing water to enter the unit.
  9. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  10. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to  any part of the RespirEasy RE-M10L medical oxygen concentrator, including  cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  11. The oxygen concentrator should be set to use in an environment without dust, corruption or toxicological harm gas.
  12. Do not place the oxygen concentrator in surroundings where its airflow is obstructed.
  13. Do not place items on top of the concentrator.
  14. This unit is not a life-support device , and in certain circumstances oxygen therapy can be hazardous, it is suggested that if any patient who needs oxygen treatment, please follow doctor's advice to choose the right flow and period for oxygen before using the oxygen concentrator.
  15. The device was expected to use in a health care environment and only trained Medical staff can operate the device
  16. In the event of an alarm, you observe your oxygen concentrator not working properly, or if you feel discomfort, consult your Equipment Provider and /or your physician immediately.
  17. Use only voltage specified on rating label.
  18. This device manufactures high concentration oxygen, which promotes rapid burning. Keep oxygen concentrator far away from open flames and no smoking around the patient.
  19. Do not leave a nasal oxygen cannula under bed coverings or chair cushions. If the unit is turned on without use, the oxygen will help the flammable material get fire. :Use no lubricants, grease, or petroleum-based products on or near your oxygen concentrator.
  20. Electrical shock hazard. Do not remove covers while the unit is plugged in. Only your Equipment Provider or a qualified service technician should remove the covers or service the unit.
  21. Care should be taken to prevent the unit from getting wet or allowing water to enter the unit.
  22. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  23. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to  any part of the RespirEasy RE-M10L medical oxygen concentrator, including  cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  24. The oxygen concentrator should be set to use in an environment without dust, corruption or toxicological harm gas.
  25. Do not place the oxygen concentrator in surroundings where its airflow is obstructed. :Do not place items on top of the concentrator.
  26. Always place the concentrator on a hard surface. Never place the concentrator on a surface such as bed or couch, where the concentrator may tip or fall.
  27. NEVER leave the concentrator unattended when plugged in.
  28. Ensure the bottom smooth exhaustion during operating, or else the oxygen concentrator will be over-heated.
  29. 5 minutes are needed from oxygen concentrator from warming up to reach regular function and nominal performance.

                                                                 3

NOTE: If oxygen does not seem to flow, first verify that the flowmeter ball is registering a flow. Then, place the tip of the cannula into a glass of water; if bubbles come out of the cannula, oxygen is flowing. If bubbles do not appear, turn off the oxygen concentrator immediately and refer to Troubleshooting.
NOTE: There is never a danger of depleting the oxygen in a room when you use your oxygen concentrator.

Radio Frequency Interference

Most electronic equipment is influenced by Radio Frequency Interference (RFI). When there is strong electromagnetic interference, maybe the LCD will be slightly affected, but the oxygen concentrator is still running. ALWAYS exercise CAUTION with regard to the use of portable communications equipment in the area around such equipment.

Requirement of environment protection

The materials used in the system won't create environment hazard. The packing materials of the system are recyclable, and they must be collected and disposed according to the related regulation in the country or region where the package of the system or its accessories is opened. The nasal oxygen tube is made of medical PVC, and if it is thrown away, it could not be bio-degradable, so it will cause the pollution. Any material of the system, that may cause pollution in the environment, must be collected disposed strictly complied with the local rules and requirements.

Contraindication
None.

4 Product  introduction

RespirEasy RE-M10L medical oxygen concentrator is a device that extracts oxygen
from atmospheric air.

It will typically be an electrically-powered molecular sieve (artificial zeolite) used to separate nitrogen from ambient air.

It could be applied widely in the hospitals at all different level, clinics, health centers and family nursing, health care for the old person, mental workers and students, etc..

 Whether the user is professional or not, &nbsp;the users could operate the oxygen concentrator by themselves after reading this user's manual.

 The oxygen concentrator can supply 1-2 patients simultaneously, with steady oxygen flowing out, safe and reliable, low cost, adjustable flow.

The key parts of the concentrator adopt anti-tiring and anti-aging design, and the planned life of the whole concentrator reaches up to 20,000 hours or 5 years, whichever comes first.

                                                                    4

5 Operation conditions and Environment

  1. Ambient temperature: 10 ℃-40 ℃
  2. Relative humidity: 30%-85%
  3. Air pressure: 700 hPa-1060 hPa
  4. Altitude: Up to 2286m without degradation; Consult your equipment provider for further information regarding to 2286m to 4000m
  5. No corrosive gas and strong magnetic field around.

6 Scope of application:

  1. For Medical Use: Medical oxygen supplied by the concentrator is beneficial to cure the disease or heart and blood vessel system, chronic pulmonary system, the brain and blood vessel system, chronic pulmonary tuberculosis, and other oxygen lacking symptoms, etc.
  2. For Health care: Medical oxygen can be used for athletics and intellectuals and brainworkers, etc. to eliminate fatigue and also suit for the departments of health care, sanatorium, healthy, plateau military camps and hotels and other places where need oxygen.

7  Technical Parameters

Model

RE-M5L

RE-M5LW

RE-M10L

Rated  power(VA)

480

540

880

Operation voltage (V/Hz)

AC230/50

Oxygen  flow(L/min)

0.5-5

0.5-5

0.5-10

Oxygen concentration (%)

93%±3%

Outlet pressure (Mpa)

0.04—0.07

Alarm

Power failure; low&high pressure ; low purity; temperature

Noise(dB(A))

≤43

≤45

≤52

                                                                                5

 

Large LCD display

Pressure digital(accuracy:0.001MPa);

accumulating timing(range:0-99999hours);

present timing(accuracy:1 minute);

presetting timing(accuracy:1 minute)

temperature digital(accuracy:0.1);

Electrical category:

Class II, Type BF

Net Weight (kg)

16

22

27

Dimension(mm)

280(deep)*348(width)*

510(height)

300(deep)*360(width)*

600(height)

365(deep)"375(width)*

600(height)

Low purity alarm (OCSI)

When oxygen purity is ≥85%, the blue lamp is on, when

oxygen purity is <85%, red lamp is on, indicating low purity Accuracy:±3%

Mains Fuse

F5AL/250V

F6.3AL/250V

F6.3AL/250V

8 Structures  and  Functions

 

                          Figure 1                                                                  Figure 2

 

                            Figure 3                                                             Figure 4

                                                                          6

           

                            Figure 5                                                             Figure 6

Figure 1 and  Figure 2 show the model RE-M10L;  
Figure 3 and  Figure 4 show the model RE-M5L; 
Figure 5 and  Figure 6 show the model RE-M5LW.

8.1  Indicating  Lamp

8.1.1 Total 8 indicating lamps and their indication for the model, such as RE-M10L  which is shown as Figure 1, are as follows:

  1. P.O:  power switch  (green  lamp)
  2. P.F.:  power  failure(red  lamp)
  3. H.P::  high  pressure(red  lamp)
  4. L.P:  low  pressure(yellow  lamp)
  5. H.T.:  over  heated  temperature(red  lamp)
  6. H.O₂:  oxygen  purity  is≥85%,(blue  lamp)(Accuracy:±3%)
  7. M.O₂:  no  use
  8. L.O..oxygen  purity  is<85%,(red  lamp)(Accuracy:±3%)

8.1.2 Total 6 indicating lamps and their indication for the model, such as JRE-M5L or RE-M5LW which is shown as Figure 3 or Figure 5, are as follows:

  1. P.O.:  power  switch  (green  lamp)
  2. P.F::  power  failure(red  lamp)
  3. L.P::  low  pressure(yellow  lamp)
  4. H.P./H.T:  high  pressure/  over  heated  temperature  (red  lamp)
  5. H.O:  oxygen  purity  is  ≥85%,(blue  lamp)(Accuracy:±3%)
  6. L.O:oxygen purity is <85%,(red lamp)(Accuracy:±3%)

8.2  Power  switch
8.3  Oxygen  flow  meter
The location of float in the oxygen flow meter shows the outlet oxygen flow (L/min.).

                                                               7

8.4 Knob of oxygen flow meter switch

  • It adjusts and controls the outlet oxygen flow.
  • Do not Switch it over-forced, or else it is easy to damage the valve core. Switch it
    counterclockwise to turn on, clockwise to turn off.

8.5 Outlet for Atomization (optional)
8.6 Intake air filter

  • Once need to replace, refer to 11.2 on page12

8.7 Storage cabinet(as shown in Figure 2⑦) It is used for storage oxygen cannula
8.8 LCD display( Liquid crystal display )(as shown in Figure 1 ⑧ or Figure 3 ⑦ or Figure 5⑦)

  • It can display some status of during operation of the oxygen concentrator, refer to
    7.on page 5
  • When starting the oxygen concentrator, the LCD screen is lighted, and it will return to screen saver mode in 15 minutes. But if you press the right key during working, the screen will be lighted again.

8.9 Timing buttons(as shown in Figure 1 ⑨ or Figure 3 ⑧ or Figure 5 ⑧)

  • The two buttons are used for timing adjustment, and each press of the left button(△) or(田) will increase timing by 10min, the max timing is 40 hours. And each press of the right button( ▽) or ( ⊙) will decrease timing by 10min. When the right button( ▽) or (θ) is pressed to reduce timing till"0", the oxygen concentrator will turn offautomatically.

8.10 Humidifier(as shown in Figure 1 ⑩ or Figure 3 ⑨ or Figure 5 ⑨)

  • Humidifier which is used for humidifying oxygen and preventing throat and nasal
    mucosa stimulated by dry oxygen and dry hard sputum difficult to spit out.

8.11 Knob of Atomizer switch (optional)(as shown in Figure 1 ⑪ or Figure 3 ⑩ or Figure 5⑩)
8.12 Rating label (as shown in Figure 2 ⑪ or Figure 4 ⑪ or Figure 6 ⑪)
8.13 Appliance inlet (as shown in Figure 2 or Figure 4 ⑩ or Figure 6⑫)
8.14 Holder(as shown in Figure 2 @ or Figure 4 B or Figure 6 B)

9 Operation  instructions

  • The plug is the disconnection device of the oxygen concentrator, when the plug is pulled, there is no power supply. In order to pull the plug easily , be certain to place the unit where all sides are at least 30 cm away from walls, draperies, furniture, or other obstructions. Do not place the unit in a confined area.

                                                                        8

  • Do not turn on or off frequently. To restart the oxygen concentrator after turning off, no less than 5 minutes are necessary (namely, exhaust internal gas of the oxygen  concentrator completely, for if air compressor turns on with pressure, its life will be shortened)

9.1 If used with a humidifier, unscrew the flask from the humidifier in clockwise direction, pour in proper distilled water or cold boiled water within the scale between the max line and the min line(see 11.4), then re-connect the top cover to the humidifier bottle, as shown in Figure 7. The humidifier used with RE-M5L/RE-M5LW/RE-M10L must comply with the general requirement of the 93/42/EEC European                                           Figure 7                          Directive as appropriate.

9.2. connect the nasal oxygen canula to the
humidifier outlet nozzle or to the concentrator
outlet if a humidifier has not been prescribed.
Then set the nasal oxygen cannula over
patient's ears, insert the nasal oxygen cannula
into patient's nostrils to absorb oxygen; The
nasal oxygen canula should be limited to 20
meters long, in order to ensure that the oxygen
flow rate remains within specification values.
The best absorbing time for health care keeps
                               Figure 8                          40-50 minutes per time, absorbing time for
                                                                      medical treatment shall be followed doctor's
                                                                      advice, as shown in Figure 8.


9.3. Insert the power plug plug into the electrical outlet of the correct voltage and frequency as defined in 7(Technical Parameters)on page 5, and the power connector connected with the
appliance inlet of the oxygen concentrator,                                       Figure 9                          then set the l/O power switch to the"!"                                                                                    position  to turn the unit on, at the same time                                                                        the O.P.lampwill light(8.1a).

                                                                   9

 

9.4. To set the flow of supplemental oxygen, turn
the knob of oxygen flow meter switch left or
right until the ball inside the flowmeter centers
on the flow line number recommended oxygen
absorbing flow.(counterclockwise—on,
clockwise—off).

 

 

                           Figure 10

Flow value:

  • 10 position flow value from 0.5~5L/min on flowmeter as shown in figure 10. 
  • The maximum recommended flow: 5L/min.
  • In compliance with the ISO 80601-2-69 standard, the flow supplied is equal to the flow set on the flowmeter, accurate to within 10% or 200ml/min, whichever is the larger of the two.
  • The variation of the maximum recommended flow does not exceed 10% of the indicated value when a back pressure of 7kPa is applied to the output of the device. The maximum.
  • outlet pressure is 70kPa。

Oxygen Concentration:
-at 2L/min:>90%
-at 5L/min: 93%(±3%)
:lt is very important to select only the prescribed level of oxygen. Change the flow selection only under the guidance of your physician.

 

    9.5.When finished the absorbing, set the I/O      power switch to the "O" position to turn off      the unit, if there is discontinuous use, please      unplug the power plug, as shown in                    Figure 11.

 

 

                         Figure 11             

9.6. If the patient needs timing oxygen absorbing, please refer to 8.9 on page 8

Atomization  operation  methods (for optional)

The atomization function of the concentrator is applicable to help to cure the sufferers of chronic pulmonary tuberculosis and respiratory system, etc. A:Atomization treatment must be under advice and suggestion of your physician. A:Atomization treatment must be under advice and suggestion of your physician. Using distilled water to do the atomization for several seconds after each operation may lighten the crystallization caused by medical solution.

                                                                   10

lf atomizing cannot work then open the cover of the bottle and add clean water in
small amount. Rotate the white ball which lies in the bottle with the gas resource connected and select the proper angle to gain a better atomization.

  • Open the cover of medical cup, and add atomization remedy that needed, then close the cover.
  • Connect the joint of atomization nozzle (or mask) with the cover of medical cup, and then connect the other end of atomizer connection tube with the atomization outlet  and turn on atomizer.
  • Turn on the power of oxygen concentrator, and shut up flow meter, then it is ready for atomization  treatment.
  • Do clean the atomization devices followed by the instructions of the atomization devices.

10 Alarms-Safety devices

10.1  Alarms

  • Power failure  alarm: In case of a loss of mains power or when the power cord is not plugged into the wall outlet, an audible alarm is activated with red indicator on(8.1.1b  and 8.1.2b on page 7). The troubleshooting is referred to 12 on page 14.
  • low&high pressure alarm: There is a pressure sensor on the main board to check the system pressure, when the pressure is lower than 0.1Mpa,there is an audible
    alarm with yellow indicator on(8.1.1d and 8.1.2c on page 7) and the oxygen
    concentrator is still working; When the pressure is higher than 0.23Mpa,there is an audible alarm with red indicator on( 8.1.1c and 8.1.2d on page 7) and the oxygen  concentrator is stopped. The troubleshooting is referred to 12 on page 14.
  • Low oxygen concentration alarm(OCSI): The oxygen concentration will rise to the normal level in five minutes of operation. When oxygen purity is ≥85%, the blue
    lamp( 8.1.1f and 8.1.2e on page 7) is on, when oxygen purity is <85%, red lamp is  on for audible alarm, indicating low purity(8.1.1h and 8.1.2f on page 7). Refer to the troubleshooting on page 14. Or call your supplier to service the device
  • Temperature alarm:  There is a temperature sensor on the main board to check the internal temperature, when the temperature is higher than 50°C in the oxygen
    concentrator, there is an audible alarm with red indicator on(see H.T. on the lamp) and  the  oxygen  concentrator  will  be  stopped.  The  troubleshooting  is  referred  to  12  on page  14.

                                                                  11

10.2 Safety devices
a. Compressor motor:
       Thermal  safety  is  ensured  by  a  thermal  switch  situated  in  the  motor                        winding(145±5℃).
b.Safety  valve:
This  is  fitted  on  the  compressor  outlet  and  is  calibrated  to  2.5  bar(250kPa).

11  Maintenance

  • Disconnect  the  power  cord  from  the  electrical  outlet  before  you  clean  the  cabinet.
  • Do  not  operate  the  concentrator  without  the  filters  installed,  or  while  filters  are  wet.

These  actions  could  permanently  damage  the  concentrator.

NOTE: If  legally  binding  regulations  govern  the  installation,  service  and/or  the
operation  of  the  product,  it  is  the  responsibility  for  the  operator  to  observe  and  follow these  regulations.
NOTE: Modifying  the  product  is  not  permitted.

11.1.  Clean  the  whole  body:  In the condition of power off,  make a clean for the outside body  by soft towel with  little  mild  household  cleaner  ,  and then wipe  it  up with  dry towel,  once  or  twice  per  month.

           

                                                            Figure  12

11.2.Clean  air  filter:  It  is  a  critical step for daily  maintenance to  clean  intake  air filter an interval  about  300  hours.
Detail  steps:  remove  the  two  intake  air filters  on  both  sides  of the  body,  clean  them with  mild  household  cleaner  and  clean  it  with  clean  water  completely,  get  ride  of  extra water and  dry  up  naturally, finally  set  back  after dry  up,  as  shown  in  Figure  13.

                                                             12

                                                                   Figure   13

11.3.  Clean  secondary  filter
Method  I:  Clean  secondary  filter  an  interval  about  800  hours,  open  the  top  cover  and  take out  storage  case,  Rotate  the  air  filter  with  counterclockwise  direction  as  Figure  14,take down  the  filter  cloth,  then  clean  it  with  detergent,  and  then  clean  it  out  with  clean  water completely,  get  rid  of the  extra water,  and  dry  it  naturally, finally  set  back  after dry  up.

                                                                 Figure  14

Method  ll:  First,  remove  the  silicone  tube  connected  with  oxygen  output,  take  down  the humidification  bottle,  Rotate  the  air  filter  with  counterclockwise  direction  as  Figure  15,take down  the  fiter  cloth,  clean  with  detergent,  wash  by  clean  water,  wipe  off  the  water.  and make it to be dry in air naturally ,then install back to the Air filter.

                                                                         13

                             

                                                                       Figure  15

11.4.  Clean  the  humidifier (if the humidifier is prescribed by a physician)
Daily:
- Empty the water from the humidifier.
- Rinse the humidifier flask under running water.  -Fill humidifier up to the mask with distilled water.
Regularly:
-Disinfect the  humidifier  parts  by  immersing them  in  a  disinfectant  solution(in  general, we recommend using water containing a small amount of chlorine bleach).
- Rinse and dry.
-Check that the humidifier lid seal is in good condition.

11.5.  Clean  Oxygen tubing  and  nasal  cannula

  •  Follow  the  manufacturer's  instructions

       11.6.  Replacement  of fuse  tube
Remove the cover of fuse, which is in the
appliance inlet, dismantle the fuse tube off by small screwdriver. Close the cover of fuse
after fuse tube is replaced.
The other fuse tube is located at the intake of
internal power line; the method of
replacement is the same with that above.           

                              Figure16   


12  Troubleshooting

If your concentrator fails to operate properly, please refer to the troubleshooting chart on the following pages for probable causes and solutions.  If problems with the equipment continue,  please  contact  your  Equipment  Provider.
NOTE: If the unit has not been used for an extended time period, it needs to operate for several  minutes  before  power failure  alarm  can  become  activated.

                                                                    14

No.

Trouble

Causes

Solution

 

 

1

 

No operation after

power connected and the P.F. lamp is light

with audible alarm

1.No  connection

between circuit of

oxygen  concentrator

and  power

2. Circuit of fuse

protector broken.

3. There is no power supply

 

1. Check out whether switch, plug, power line in good

connection.

2.Replace the fuse protector and find the cause

 

 

 

2

 

 

No oxygen out or tiny outtake flow

1. Folded inside

oxygen tube, no

smooth  outtake 2.Filter clogged, no

smooth  intake

3.The cover of

dampen bottle leaking

1.Connect the oxygen tube

again

2. Clean the filter

3.Take off the cover, screw well the cover, block the outtake

by thumb after turning on, and there will some sound from

the humidifier after 5 second around (the safety valve of humidifier turns on)

 

3

 

No exhaust sound

1.Air controller cannot work

2. Electrical control board cannot work

1.Have air control valve

replaced

2.Have electric control board replaced

 

4

 

Too noisy exhaustion

1. The joint of

exhaustion muffler

fallen off

2. Exhaustion muffler broken

1.Connect the joint well

2.Have the muffler replaced

 

5

The  oxygen

concentrator is working

but the L.P. lamp is

light with audible alarm.

The system pressure is too low

Check every gas circuit

connectors with soapy water

whether there are air leakage.

 

 

6

The  oxygen

concentrator is stopped and the H.T. lamp is

light with audible alarm.

 

The temperature in the oxygen concentrator is too high.

1. Check the fan's connector on the main board whether it is bad contact.

2. turn off the oxygen

concentrator and consult your Equipment  Provider.

 

7

The  oxygen

concentrator is stopped and the H.P. lamp is

ight with audible alarm.

The system pressure is too high.

Turn off the oxygen concentrator and consult your Equipment

Provider.

 

 

8

The  oxygen

concentrator is working

but the L.O  lamp is light.

 

Oxygen  concentration is too low

1. Check every gas circuit

connectors with soapy water whether there are air leakage

2. Turn off the oxygen

concentrator and consult your Equipment  Provider.

                                                                         15

13  Information  on  Electromagnetic  compatibility

The RE-M5L/RE-M5LW/RE-M10L needs special precautions regarding and needs to be installed and put into service according to the EMC information provided in the accompanying  documents;
Portable and mobile RF communications equipment can affect the RE-M5L/RE-M5LW/RE-M10L. 
All cables and maximum length of cables, Transducers and other accessories with which the manufacturer of the RE-M5L/RE-M5LW/RE-M10L claims compliance with the requirements, Accessories that do not affect compliance with the requirements of these
sub clauses need not be listed. Accessories, transducers and cables may be specified either generically or specifically.
NOTE:
Transducers and cables sold by the manufacturer of the RE-M5L/RE-M5LW/RE-M10L as replacement parts for internal components need not be listed.
The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by the manufacturer of The RE-M5L/RE-M5LW/RE-M10L as replacement parts for internal components, may result in increased emissions or decreased immunity of The RE-M5L/RE-M5LW/RE-M10L.

Guidance and manufacturer's declaration - electromagnetic emissions

The RE-M5L/RE-M5LW/RE-M10L is intended for use in the electromagnetic environment

specified below. The customer or the user of the RE-M5L/RE-M5LW/RE-M10L should assure that it is used in such an environment.

Emissions test Compliance Electromagnetic    environment-     guidance
RF emissions
CISPR 11
Group 1 The RE-M5L/RE-M5LW/RE-M10L uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The RE-M5L/RE-M5LW/RE-M10L is suitable for use in all establishments other than domestic, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

However, the following warning applies: WARNING: The RE-M5L/RE-M5LW/RE-M10L is for use by healthcare professionals only.

This device/system may cause radio interference or interfere with the operation of nearby equipment. Mitigation measures may be required, such as reorienting or relocating the RE-M5L/RE-M5LW/RE-M10L or shielding the location.
Harmonic
emissions
IEC 61000-3-2
Class A
Yoltage
fluctuations/
flicker emissions IEC 61000-3-3
Complies

16

Guidance and manufacturer's declaration - electromagnetic immunity

The RE-M5L/RE-M5LW/RE-M10L is intended for use in the electromagnetic environment

specified below. The customer or the user of the RE-M5L/RE-M5LW/RE-M10L should assure that it is used in such an environment.

IMMUNITY

test

IEC 60601

test level

Compliance level

Electromagnetic

environment- guidance

Electrostatic

discharge

(ESD)

IEC 61000-4-2

 

± 8 kV contact ±15 kV air

 

± 8 kV contact

±15 kV air

Floors should be wood,

concrete or ceramic tile. If

floors are covered with synthetic material, the

relative humidity should be at least 30 %

Electrical fast transient

burst IEC 61000-4-4

±2 kV for power supply lines

±1 kV for

input/output

Lines

± 2 kV for power supply lines

±1 kV for

input/output

Lines

Mains power quality should

be that of a typical

commercial or hospital environment.

Surge

IEC  61000-4-5

± 1kV line(s) to

line(s)

±2 kV line(s)to earth

± 1kV line(s)to

line(s)

±2 kV line(s) to earth

Mains power quality should be that of a typical

commercial or hospital environment.

Voltage dips,

short

interruptions

and

voltage

variations

on power

supply

input lines

IEC 61000-4-

11

<5% UT

(>95 % dip in UT) for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70% UT

(30 % dip in UT) for 25 cycles

<5% UT

(>95 % dip in UT) for 5 s

<5 % UT

(>95 % dip in UT) for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

<5% UT

(>95 % dip in UT) for 5 s

Mains power quality should be that of a typical

commercial or hospital

environment. If the user of the

 RE-M5L/RE-M5LW/RE-M10L  requires

continued operation during power

 mains interruptions, it is recommended 

that the RE-M5L/RE-M5LW/RE-M10L

be powered from an

uninterruptible power

supply or a battery.

 

 

 

Power

frequency

(50/60 Hz)

magnetic field IEC 61000-4-8

 

 

 

 

3 A/m

Not applicable

Note: The RE-M5L/RE-M5LW/RE-M10L does not contain components

susceptible to magnetic fields, such as Hall elements or

magnetic field

sensors. Therefore, the EUT is deemed

to meet the

requirement without actual testing.

 

 

 

Power frequency magnetic fields should

be at levels characteristic of a typical

 location in a typical commercial or 

hospital environment.

17

Guidance and manufacturer's declaration - electromagnetic immunity

The RE-M5L/RE-M5LW/RE-M10L is intended for use in the electromagnetic environment specified below. The customer or the user of the RE-M5L/RE-M5LW/RE-M10L should assure that it is used in such an electromagnetic environment.

IMMUNITY

test

IEC 60601

test level

Compliance

level

Electromagnetic environment - guidance

 

 

 

 

 

 

 

 

Conducted RF IEC 61000-4-6

 

 

 

Radiated RF

IEC 61000-4-3

 

 

 

 

 

3V

0.15MHz-80

MHz

6V in ISM

bands

between

0.15MHz and 80MHz

80% AM at

1kHz

3 V/m

80 MHz to

2.7 GHz

80% AM at 1kHz

 

 

 

 

 

3V

0.15MHz-80

MHz

6V in ISM

bands

between

0.15MHz and 80MHz

80% AM at

1kHz

3 V/m

80 MHz to 2.7 GHz

80% AM at

1kHz

Portable and mobile RF

communications equipment should be  used no closer to any part of the RE-M5L/RE-M5LW/RE-M10L,including cables, than the recommended

separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

 

 

 

d=1.17 √P d=1.17

 √P d=2.33 √P

80 MHz

800 MHz

to 800 to 2,7

MHZ

Ghz

 

 

 

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter

manufacturer and d is the

recommended separation distance in metres (m).

Field strengths from fixed RF

transmitters, as determined by an

electromagnetic site survey, should be

less than the compliance level in each frequency range."

Interference may occur in the vicinity of

equipment marked with the following

symbol: 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is

 affected by absorption and reflection from structures, objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the RE-M5L/RE-M5LW/RE-M10L is used exceeds theapplicable RF compliance level above, the RE-M5L/RE-M5LW/RE-M10L should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the RE-M5L/RE-M5LW/RE-M10L.

5 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

   18

Recommended separation distances between portable and mobile RF communications equipment and the JAY-5AWIJAY-5BW/JAY-10

The RE-M5L/RE-M5LW/RE-M10Lis intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the RE-M5L/RE-M5LW/RE-M10L can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment

(transmitters) and the RE-M5L/RE-M5LW/RE-M10L as recommended below, according to the maximum output power of the communications equipment.

Rated maximum

output power

of transmitter

W

Separation distance according to frequency of transmitter m

 

150 kHz to 80 MHz d=1.17 √P

80 MHz to 800 MHz d=1.17 √P

800 MHz to 2,7 GHz d=1.17 √P

0.01

0.12

0.12

0.07

0.1

0.37

0.37

0.22

1

1.17

1.17

0.70

10

3.69

3.69

2.21

100

11.67

11.67

7.00

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable

 to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and

reflection from structures, objects and people.

14 Accessories

1. Your concentrator includes the following components:

  •  Intake air filter (two pieces, part number:G-01)
  • Secondary filter (one piece, part number:G-02)

The concentrator comes with two air filters and one secondary filter already installed.
2. Please use the parts mentioned in this chapter, if the use of other parts can
degrade minimum safety and performance.
3. Please choose the suitable humidifier and Nasal oxygen cannula, they must:

  • be oxygen compatible,
  • be biocompatible,
  • include a means to prevent the propagation of fire and accord with requirements of ISO 80601-2-69

                                                                     18

15  Condition for transportation and storage

  • Environment temperature scale:-40~55 ℃
  • Comparative humidity scale:≤95%
  • Air  pressure  scale:  700-1060  hpa

16  Quality  Warranty

  • Warranty  for  whole  unit:15  months
  • Warranty  for  magnetic  valve:24  months
  • Warranty  for  compressor:24  months

17  Contact  us

20

Back to blog

Leave a comment

Please note, comments need to be approved before they are published.

About RESPIREASY: Leanding Manufacturer of Oxygen Concentrator

About RespirEasy: Breathe Freely, Live Fully  

Slogan: Easy Respiratory Care, Smart Oxygen Solutions  

RespirEasy has been rooted in the oxygen therapy equipment industry for 15 years, and we have always prioritized "providing reliable products and thoughtful services". Our product portfolio covers 1-60L full-range oxygen concentrators, along with original accessories such as filters and humidifiers. Whether for daily home use, portable on-the-go needs, or professional medical settings, we can meet the oxygen therapy requirements of different users, and have won the recognition and trust of over 400,000 users to date.  

 

Solid Manufacturing, Quality First  

Our factory is located in Shenzhen China, covering an area of 3,500 square meters and holding professional certifications. We have also established a comprehensive advanced technical management system, with a stable annual production capacity of 2 million units. Every oxygen concentrator is manufactured in strict compliance with high European and American standards — from the screening and testing of components, to the control of production processes, and the inspection of finished products before delivery, no link is overlooked. This is to ensure that users worldwide can use our products with confidence, and that the quality, safety and performance of the products can stand the test of long-term use.  


Cross-Border R&D, Innovation-Driven  

Innovation is not a slogan, but our core competitiveness. We have established professional R&D teams in Germany, renowned for engineering precision, and Shenzhen, a global high-tech hub. The two teams together bring together more than 30 senior scientific research experts. Through cross-border collaboration and technical complementarity, we continuously promote the iteration and upgrading of oxygen therapy technology. Up to now, the company has accumulated more than 100 patents and 17 software copyrights. All R&D and innovation efforts are closely focused on the actual pain points of users, aiming to create exclusive oxygen therapy solutions that better fit real-world scenarios.  

 

 

Multi-Scenario Models, Customized to Needs  

  • Portable Models  

The all-new 2025 lineup includes a 7L handheld model (only 3.14 lbs) and a 5L continuous flow backpack model. Both are equipped with lithium batteries and dedicated carrying bags, supporting free switching between three power supply modes: car, home, and battery. This completely eliminates users' worries about oxygen therapy while traveling, ensuring stable oxygen supply anytime, anywhere.  

  • Home/Medical Models  

The 1-9L adjustable models come with a built-in nebulization function, catering to both oxygen therapy and nebulization needs. The 10-60L high-power medical-grade devices maintain an oxygen concentration stably at the medical standard of 93%-95%, and operate with low noise, providing a more comfortable experience for long-term users.  

  • Original Oxygen Accessories  

We supply a full range of original accessories, including DEDAKJ water tanks, water humidifiers, oxygen filters, oxygen tubes, and nasal cannulas. This avoids equipment malfunctions caused by incompatibility issues with non-original accessories, ensuring the long-term stable operation of oxygen concentrators.  

 

Core Advantages, A Worry-Free Choice  

- Fast Delivery: With local warehouses in the US, Mexico, and Europe, orders can be delivered to users within 3-4 days after placement, no long wait required.  

- Quality Warranty & 24 Hours Support: All products come with a 1-year warranty, and we also offer a 180-day price guarantee. Our professional after-sales team responds promptly, ensuring that any user questions are addressed in a timely manner.  

  • Charitable Support: We understand the importance of high-quality oxygen therapy equipment for some families. Therefore, we specially provide 10 discounted units every month to help low-income families access reliable oxygen therapy products, ensuring that good equipment truly benefits more people in need.

 

Click the link below to buy Oxygen Concentrator:

>>Buy 3-9 liters mini handy portable oxygenn concentrators

>>Buy 1-9 liters home personnal use oxygen concentrators

>>Buy 10-60 liters powerful medical oxygen concentrators

 

Have a question?

If we still haven't answered your question, you can contact us at the following and we will get back to you as soon as possible.

Contact Us:

Email: support@airoxygenconcentrator.com